AWD-01C
Automatic Open Cup Flash Point Tester
ASTM D92
AWD-01C Automatic Open Cup Flash Point Tester ASTM D92 is a sophisticated device meticulously engineered to align with the stringent requirements outlined in GB/T 3536-2008 "Method for Determination of Flash point and Ignition Point of Petroleum Products (Cleveland Open Cup Method)" and GB/T267 open flash point standards of the People's Republic of China. AWD-01C cleveland open cup apparatus ASTM D92 is specifically tailored for the Cleveland Open Cup Method a widely recognized procedure for determining the flash points of petroleum products that have had fuel oil removed and possess open flash points below 79 °C. The AWD-01C Flash Point Testing Cleveland Open Cup Apparatus ASTM D92's applicability spans across various sectors including railway, aviation, electric power, petroleum industry, colleges and universities, scientific research institutes, measurement and testing departments, and other units that require precise detection and testing of the flash point and ignition point of petroleum products.
The significance of the AWD-01C cleveland open cup apparatus ASTM D92 lies in its adherence to the Cleveland Open Cup Method, a method that is also the foundation of the ASTM D92 standard. The
ASTM D92 Flash Point Testing Cleveland Open Cup Method titled "Standard Test Method for Flash and Fire Points by Cleveland Open Cup Tester," provides a systematic approach to measuring the flash point and fire point of petroleum products.using the Cleveland Open Cup Apparatus.
The flash point, as defined by ASTM D92 standard, is the lowest temperature at which the application of an ignition source causes the vapors of a specimen to ignite under specified conditions. The fire point, conversely, is the lowest temperature at which the vapors of a specimen sustain burning for a minimum of 5 seconds under the same conditions.
Flash point testing is a critical safety procedure that assesses the potential hazards associated with the storage and handling of flammable liquids. The flash point is an indicator of the temperature at which a liquid's vapor can form a flammable mixture with air, which is essential for determining the safe storage and transportation conditions of these materials. The Cleveland Open Cup method, in particular, is favored for its ability to simulate real-world conditions where the vapors produced by heating the test liquid are allowed to escape to the atmosphere, thus providing a more accurate representation of the substance's behavior in open environments.
The AWD-01C ASTM D92's design incorporates the Cleveland Open Cup apparatus, which is a key component in conducting flash point testing according to the ASTM D92 standard. This apparatus allows for the controlled heating of a sample and the introduction of a test flame to determine the flash point. The AWD-01C Automatic Open Cup Flash Point Tester ASTM D92's operation is based on the principle of measuring the properties of a sample in response to heat and a test flame under controlled conditions, ensuring that the flash point measures the tendency to form a flammable mixture with air while the fire point indicates the tendency of sustained burning.
The AWD-01C cleveland open cup apparatus ASTM D92 is not only designed for precision but also for versatility, making it suitable for a wide range of applications. Its ability to determine the flash points of petroleum products with fuel oil removed and open flash points below 79 °C makes AWD-01C Automatic Open Cup Flash Point Tester ASTM D92 an indispensable tool in industries where the safety and handling of flammable materials are of paramount importance. AWD-01C Automatic Open Cup Flash Point Tester ASTM D92's wide range of applications is a testament to its robust design and adherence to international standards.
AWD-01C ASTM D92's alignment with the Cleveland Open Cup method and its compliance with the ASTM D92 standard make it a valuable asset in the field of flash point testing. Its ability to provide precise measurements of the flash point and ignition point of petroleum products is crucial for ensuring safety in various industries. The device's wide range of applications, from railway and aviation to scientific research and measurement testing, underscores its importance in maintaining safety standards and regulatory compliance.
In conclusion, the AWD-01C Automatic Open Cup Flash Point Tester is a state-of-the-art device that offers unparalleled accuracy and reliability in flash point testing. Its design and functionality are in strict accordance with the requirements specified in GB/T 3536-2008 and GB/T267 standards, making it an essential tool for determining the flash points of petroleum products. AWD-01C Automatic Open Cup Flash Point Tester ASTM D92's adherence to the Cleveland Open Cup method and its compliance with the ASTM D92 standard ensure that it provides precise and consistent measurements, which are vital for safety and regulatory compliance across various industries.
cleveland open cup apparatus flash point test ASTM D92
cleveland open cup apparatus flash point test ASTM D92
cleveland open cup apparatus flash point test ASTM D92
Functional Features of cleveland open cup apparatus flash point test ASTM D92 Flash Point Testing Cleveland Open Cup Method
1. Automatic open cup flash point tester ASTM D92 used to determine the open cup flash point value of petroleum products.
2. 8-inch pure industrial true color touch screen, English and Chinese interface can be switched at will, Windows CE embedded operating system.
2. With the instrument self-diagnosis function to check whether there is a fault and display its status.
3. Automatically correct the influence of atmospheric pressure on the test and calculate the correction value.
4. Standard RS485 interface, also can be connected to USB2.0, (optional) can set the instrument parameters and upload data to the computer. Can store up to 120 sets of historical data.
5. The ignition mode is selected manually and automatically.
6. Detection, cover opening, ignition, printing, automatic completion.
7. Using the original imported Renesas microcontroller. PT1000 high precision temperature sensor imported from Germany.
8. With the function of preventing fire overtemperature and automatically stopping work.
9. The instrument is equipped with ignition cover to automatically extinguish the ignition flame.
10. Built-in high-precision flash point detection sensor, detection time up to milliseconds.
11. Simulation tracking shows the function curve of temperature rise and test time.
Automatic Open Cup Flash Point Tester ASTM D92
Automatic Open Cup Flash Point Tester ASTM D92
Technical Parameters of cleveland open cup apparatus flash point test ASTM D92 Flash Point Testing Cleveland Open Cup Method
Power Supply: AC220V 50HZ
Temperature Measurement Range: room temperature ~ 400℃
Repeatability: ≤8℃
Resolution: 0.1℃
Accuracy: 0.5%
Base Parameters: temperature rise rate: in line with GB/T3536 standard
Ignition Mode: electronic ignition,gas flame
Ambient Temperature: (10 ~ 40)℃
Relative Humidity: ≤80%
Power Consumption: ≤ 500W
11. Standard : ASTM D92